100-130V Low Voltage Controller for Four-Gang
or Larger Masonry Box Applications

The LC-040 enables users to safely operate the high-voltage lighting found in hospital and nursing home rooms with their pillow speaker or bed communication side rail. It’s tested and approved for use with all Nurse Call systems that use a momentary normally open switch for auxiliaries, and is designed for applications using a four-gang or larger masonry box.

  • Can control lamps individually or sequentially
  • Can be used to operate high-voltage circuits such as lights, electric doors, and drapes
  • Includes NEC isolation barrier plates
  • No jumpers required to differentiate between one or two switch control
  • Emits no electromagnetic noise that will be detected by sensitive medical equipment
  • Isolation from line voltage via a Class II transformer
  • One year warranty
  • UL/C-UL Listed to UL 508
  • Compliant with FCC, Title 47 CFR 15 – Class B
  • RoHS compliant
Input voltage 100-130 VAC, 50/60 Hz
RMS current should not exceed 5 amps.
Load Type 277 VAC, 5 A, General Use per Load277 VAC, 5 A, Ballast per LoadMaximum total device current: 10 A
Output Voltage 100-130 VAC, 50/60 Hz 
Output Devices Two 10 Amp relays driven by low voltage
microcontroller, ESD protected
Dimming Control NA
Low Voltage Momentary Switch Requirements Dry contacts, momentary, normally open, capable of switching 5 VDC @ 0.5 mA
Lamp No restrictions
Switching circuits operation voltage 5 VDC at 2.3 mA
Power Supply Classification Isolation from line voltage via a Class 2 transformer
Operating Temp. Range 40º to 105º F (5º to 40º C)
External wires White/Black: 16 AWG stranded All other wires: 18 AWG stranded
Mounting 3.5" deep masonry box, four gang min.
Operating Temp. Range 40º to 105º F (5º to 40º C)
External Wires White and black wires: 16 AWG stranded
All other wires: 18 AWG stranded
Mounting Designed to mount into a 3.5” deep masonry box with a minimum of four gangs
Housing 058" Galvanized Zinc-Plated Steel
Dimensions 3.5” (L) x 3.5” (W) x 3.8” (D)
Warranty One Year
Regulatory Listings and Compliance • UL/C-UL Listed to UL 508;
• Compliant with FCC, Title 47 CFR 15 – Class B;